How to eliminate the high pressure alarm of the industrial ultra-low temperature freezer?
During use, if a fault is not solved in time, it will affect the use of industrial ultra-low temperature freezer. How to solve the problem of high-pressure alarm?
If the water-cooled industrial ultra-low temperature freezer presents a high-pressure alarm, it is mainly related to the cooling water system. The water-cooled industrial ultra-low temperature freezer provides circulating cooling water through a cooling tower. If you forget to open the cooling water valve during use, If the cooling water cannot be circulated to the chiller, it will show high pressure problems. In addition, check whether the cooling water tower is defective and cannot safely and stably deliver the cooling water.

The cooling water flow is too small, or the temperature is too high and the heat dissipation is not good. This problem will also occur. If the cooling water flow is small, it is necessary to check whether the pipe size of the supporting device matches the size of the industrial ultra-low temperature freezer group, or whether the water valve can be opened to a large position; if it is an air-cooled industrial ultra-low temperature freezer, we need to look at the ambient temperature of the device If it can be exceeded, the heat must be exhausted in time to ensure that the industrial ultra-low temperature freezer unit is in an air circulating environment.
Industrial ultra-low temperature freezers are used for too long, and it is easy to accumulate dirt and other debris on the pipe wall or condenser copper pipes, which will inevitably affect the refrigeration cycle system. Therefore, we regularly provide industrial ultra-low temperature freezers according to the maintenance time recommended by the manufacturer Do maintenance cleaning work (cleaning the cooling water system, cleaning the condenser, changing the filter, etc.).
Check whether the refrigerant is excessive. The industrial ultra-low temperature freezer unit will be filled with the appropriate amount of refrigerant before leaving the factory. Therefore, if there is no leakage of the refrigerant, the user does not need to charge the refrigerant. Too much refrigerant will only affect the unit’s failure. use.
Whether the opening of the shrink valve is too small, for this, everyone can gradually adjust the opening of the shrink valve to eliminate the problem.
Verwandte Empfehlungen
-
Explosion-proof high temperature and low temperature machine factory note
1219Explosion-proof high and low temperature integrated machine is the main product of LNEYA production and sales. It adopts dynamic temperature control system and fully closed pipeline design. The operation effect is relatively stable. What common pr...
Details anzeigen -
Classification and advantages and disadvantages of water-cooled chillers
1335Industrial chillers are divided into air-cooled industrial chillers and water-cooled industrial chillers. Today we will discuss the classification, advantages and disadvantages of water-cooled chillers. Advantages of water-cooled chillers: 1. Inte...
Details anzeigen -
The reason for the existence of cryogenic refrigerator in the tool parts manufacturing industry
870Cryogenic treatment is applied to worn tools, die cutter materials, and later extended to alloy steel, cemented carbide, etc., this method can change the internal structure of metal materials, thereby improving the mechanical and processing proper...
Details anzeigen -
Attention points for the use of laboratory chiller manufacturers’ equipment
885In the laboratory, a wide range ofexperimental instruments are widely used. When the experimental instruments arein operation, they are inseparable from the high-low temperature circulatingdevice refrigeration heating equipment. What should be ...
Details anzeigen