Anweisungen zur Auswahl des Volumens der kleinen Hoch- und Tieftemperatur-Schlagprüfkammer
When the test product (components, components, components or complete machine) is placed in a small high and low temperature impact test chamber for testing, in order to ensure that the atmosphere around the test product can meet the environmental test conditions specified in the test specification, the small high and low temperature impact test There should be some points of attention between the working size of the box and the outer dimensions of the tested product. The volume of the tested product should not exceed (20~35)% of the effective working space of the test box. It is recommended to use no more than 10% for products that generate heat during the test. The ratio of the windward cross-sectional area of the test product to the total area of the test chamber on the section is not more than (35 to 50)%. The distance of the outer surface of the test product from the wall of the test chamber is at least 100 to 150 mm.
After the test piece is placed in the box, the smooth passage is occupied, and the narrowing of the passage will result in an increase in the flow rate of the air flow. Accelerate the heat exchange between the gas stream and the test piece. This is inconsistent with the reproduction of environmental conditions, because in the relevant standards, the temperature flow around the test specimens in the test chamber should not exceed 1.7 m/s, so as to prevent the test specimens and the surrounding atmosphere from being produced. Heat transfer. If the volume of the test piece or the wind-breaking area is increased without restriction in the test, the airflow speed will increase beyond the wind speed specified in the test standard during the actual test, and the validity of the test results will be suspected.

The accuracy index of the environmental parameters in the working chamber of the small high and low temperature impact test chamber is the result of the test under no-load condition. Once the test piece is placed, the uniformity of the environmental parameters in the working chamber of the test box will be affected. The greater the space occupied, the more serious this effect will be.
According to the principle of heat conduction, the temperature of the gas flow near the tank wall is usually 2 to 3 °C from the center temperature of the flow field, and may reach 5 °C at the upper and lower limits of the high and low temperature. The temperature of the tank wall is different from the temperature of the flow field near the tank wall by 2~3 °C (depending on the structure and material of the tank wall). The greater the difference between the test temperature and the external atmospheric environment, the greater the above temperature difference, therefore, the distance from the tank wall The space within the distance (100 to 150 mm) is an unusable space.
The volume selection instructions above the small high and low temperature impact test chamber do not know that you understand it. I believe that users can better select the equipment after reading these.
Verwandte Empfehlungen
-
Cooling heating temperature control system SUNDI series thermal fluid filling instructions
11101. Refrigeration heating temperature control system SUNDI series equipment before the heat conduction filling, the heat transfer liquid inlet of the heating refrigeration circulator and the heat transfer liquid outlet of the upper end of the react...
Details anzeigen -
Effect of oil separator on ultra-low temperature refrigerator
1304In ultra-low temperature chillers, oil separators are an important accessory. In most cryogenic refrigerators, there are oil separators. What are the effects of oil separators on cryogenic refrigerators? Needless to say, the position of the cryoge...
Details anzeigen -
Analyse von häufigen Fehlern im Batterietestsystem für New Energy Vehicle Power
948Wenn die neue Energie-Fahrzeug-Power-Batterie-Pack Lecks, die allgemeine Fehler Manifestation Modus: das Messgerät OK Licht nicht leuchtet, fordert das Messgerät, um das Stromnetz zu überprüfen, die Hochspannungs-System Leckage Fehler, trennen Sie die Batterie-Pack und die b...
Details anzeigen -
Design and Test Technology of High Temperature Environmental Chamber Sales
12201. Temperature selection (1) The climatic zones of the region; Whether the product is exposed to the solar radiation environment near the heating device; Surrounding air, solar radiation and ways of heat transfer from heating devices to products;...