The enterprise users are often at a loss when dealing with common faults in the salt spray environmental chamber. For this reason, Wuxi LNEYA, a manufacturer of the salt spray environmental chamber, specially reorganized the related common fault treatment technologies of the salt spray environmental chamber.
Salt spray environmental chamber for sale
The salt spray environmental chamber cannot be heated
1. Heat the solid state relay, if there is no output, replace the solid state relay.
2. If the external dial has no output, replace the external dial.
3. If there is heating tube damage, check the heating tube resistance, replace the resistance if there is something abnormal.
4. Lack of water can cause heating shielding, check whether the water level is normal or not and whether the floating ball malfunctions or not.
5. If there is a malfunction of the over-temperature maintenance system, adjust the maintenance temperature value. If the fault still cannot be solved, replace the over-temperature maintenance system.
The heating temperature is abnormal
1. If the set temperature of the temperature controller of the environmental chambers for sale is too low, set the temperature controller at the required temperature.
2. If the set temperature of the safety maintenance switch of the environmental chamber is too low, set the safety maintenance switch at the required temperature.
3. If the set temperature of the temperature controller of the full bucket is too low, set the temperature controller at the required temperature.
4. If the set temperature of the safety maintenance switch of the full bucket is too low, set the safety maintenance switch at the required temperature.
The external dial doesn’t display
1. Check whether or not there is a power supply in the salt spray environmental chamber. If not, ask a professional electrician for a check.
2. Check whether or not the power switch is damaged. If damaged, replace the power switch.
3. Check whether or not the power circuit breaker is damaged. If damaged, replace the circuit breaker.
4. Check whether or not the external dial terminal has a power supply. If there is a power supply, it indicates that the external dial is damage, replace the external dial.
5. Check whether or not the power supply wiring terminal of the salt spray environmental chamber is blackened. If it is blackened, it indicates the possibility of burning out, replace the external dial.
If the failures the enterprise users are encountered are more serious than above, it is suggested to contact the salt mist environmental chamber manufacturer to solve the failure.
Verwandte Empfehlungen
-
500L reactor distributed temperature control unit system
791The 500L reactor is a relatively large-capacity reaction vessel in industrial manufacturing and pharmaceutical companies. It is usually used in pilot-scale or small- and medium-scale production stages, and widely serves multiple industries...
Details anzeigen -
Operating Principle of LNEYA Industrial Air Cooled Chillers
1323Industrial Air Cooled Chillers is a heatexchanger which USES air to cool the hot fluid. The hot fluid in the tube isheated through the exchange of heat between the tube wall, the fin and theoutside air, which is usually supplied by the ventilat...
Details anzeigen -
Wie geht man mit einer Störung des industriellen Kaltwassersatzes um?
1155Was stimmt nicht mit dem industriellen Wasserkühlsystem, wenn es nicht abkühlt? Lassen Sie es sich von LNEYA erklären: Ein Grund ist, dass die anfängliche Verwendung des Flüssigkeitskühlsystems die Temperatur nicht senken kann, oder es ist etwas mit dem ...
Details anzeigen -
Semiconductor production process and testing process temperature control unit
923LNEYA semiconductor testing process chiller
Details anzeigen